How Regenis delivers what conventional injectors block

Regenerative biostimulators are suspensions of solid microspheres. Conventional injector needles are specified for solutions. This page sets out the delivery problem that follows, and the engineering Regenis uses to address it.

The failure mode

Microspheres pack and bridge at the needle hub

Where the hub narrows into the cannula, suspended microspheres can pack and bridge across the opening

Scanning electron micrograph of an occluded needle hub
Scanning electron micrograph of an occluded needle hub. Manufacturer data, Hertz Biotechnology (Shenzhen).

Needle tips also change shape with reuse — a separate problem from hub packing, documented on the same manufacturer plates.

Scanning electron micrographs comparing a needle tip after five and after ten injections
Needle tip after five injections and after ten, with detail insets. Tip geometry changes with reuse. Manufacturer data, Hertz Biotechnology (Shenzhen).

Why it happens

Manual reconstitution does not distribute evenly

Biostimulator microspheres are solid and denser than their carrier, so they settle. How thoroughly a syringe is mixed by hand determines how evenly the dose is distributed — and that varies between operators and between passes.

X-ray of syringe contents after five passes and after ten passes
5 次混合10 次混合
X-ray of syringe contents after five mixing passes (left) and ten (right). Distribution differs visibly between the two. Manufacturer data, Hertz Biotechnology (Shenzhen).

Sensing and control

Five signals, sampled continuously, resolved in a closed loop

Sensors integrated into the handpiece report at millisecond intervals. The control model reads all five together and adjusts drive output while the injection is in progress, rather than after it.

Perception

  • Negative pressure
  • Thrust force
  • Speed
  • Current
  • Torque

Control model — on-device

Quad-core ARM CPU with a dedicated AI processing unit. Injection planning, clogging warning and risk analysis run on a real-time operating system.

Actuation

  • Power
  • Rotation direction
  • Speed

Sensors report at millisecond intervals; drive output is adjusted while the injection is in progress.

Protected design

The anti-clogging mechanism is patented

The delivery path is the subject of a granted utility-model patent held by the manufacturer

Holder
Hertz Biotechnology (Shenzhen) Co., Ltd.
Title
Microsphere injection anti-clogging device
Application number
202421307575.3
Filed
7 June 2024

Precision and interlocks

The system reports when delivery is not nominal

Nine needles share one regulated feed, so coverage is uniform across the treated area. When the drive train meets resistance outside its expected envelope, the system raises a named alarm rather than continuing silently.

Alarm states
Low suction pressure · Handle disconnected · Pusher blocked
Needle array
Nine-needle, uniform coverage
Applied part
IEC Type B
Negative pressure
Up to 90 kPa, 10 levels

In the treatment room

Built to be handled through a full clinic day

Weight, cleaning and consumable handling determine whether a device is used or left in a cupboard

Handpiece weight
212 g
Filtration
Sterile negative-pressure tubing, triple filtration
Line cleaning
One-touch vacuum line cleaning, 750 ml/min
Memory presets
Eight
Handpiece dock
Magnetic
Noise
< 52 dB

Range covered

Specified for biostimulators and conventional mesotherapy solutions

The delivery path is designed around the thicker suspensions that block conventional injectors, without excluding the lighter formulations already in routine use

Biostimulators
PDLLA · PLLA · PCL · CaHA
Mesotherapy solutions
Basic · Nutritional · Functional

Specifications

技术规格

Display
13.3" HD touchscreen
Handpiece weight
212 g
Needle array
Nine-needle, uniform coverage
Negative pressure
Up to 90 kPa, 10 levels
Memory presets
Eight
Control hardware
Quad-core ARM CPU + AI processing unit
Operating system
Real-time Linux
Line cleaning
One-touch vacuum line cleaning, 750 ml/min
Filtration
Sterile negative-pressure tubing, triple filtration
Noise
< 52 dB
Handpiece dock
Magnetic
Motor durability
1.6 million injection cycles
Cable
Rated to 100,000 connection cycles
Compliance
GB 9706.224-2021 · YY 9706.108-2021
临床光照下的肖像

联系我们

申请产品演示

请填写以下表格,我们的团队将与您联系以安排产品演示。

或直接联系我们

[email protected]

我们力争在 1–2 个工作日内回复。·标有 * 的栏位为必填。

我们仅将您提供的信息用于回复咨询,并在您提出时安排产品演示。请勿提交患者或敏感健康数据。请参阅我们的 隐私政策。

Regenis

越南由 RX Dermatech 有限公司经销

越南胡志明市西贡坊海潮街 02 号 Bitexco 金融大厦 46 层

公司注册号:0318788456·越南 B 类适用标准公示号:260000821/PCBB-HCM · 公示日期 2026年3月26日